MLS Wave Filtering in Limb-Tracking Observations
If MLS radiances are saturated during the limb-tracking observations, the weighting function does not move in the vertical plane relatively to the tangent point. Depending on its orientation with respect to wave fronts, MLS variance may have maximum response to wave perturbation in case A and minimum response in case B due to the weighting function averaging. When projected onto the horizontal plane, the weighting function averaging depends also on horizontal wavelength and viewing angle (that is defined as the angle between wave propagation and line-of-sight directions). As a result of 3D wave propagation, the instrument wave filter during limb tracking operation peaks at viewing angles of ~110o and ~250o, which deviate slightly from 90o and 270o as would expected for the situation only involving the horizontal cross-section.
Vertical cross-sectionCase A
Case B
Horizontal cross-section
![]()
Instrument wave filter as a function of viewing angle